Background: Hemolysis is the most frequent and potentially one of the most harmful causes of poor sample quality. Many strategies have been attempted for minimizing the risk of generating spurious hemolysis during blood collection, such as using the HOLDEX® single-use holder, which features a chamber that decelerates blood flow pressure before entering evacuated blood tubes. We have hence designed a study to establish whether the new VACUETTE® SAFELINK, which shares different structure but similar strategy affecting blood flow pressure, may ensure a comparable sample quality as using HOLDEX® single-use holder. Methods: A total number of 24 clinical chemistry, hematologic and hemostasis parameters were measured in paired blood samples collected from 30 healthy volunteers using either HOLDEX® single-use holder or VACUETTE® SAFELINK. Test results were compared with paired Student’s t-test, Pearson’s correlation and Bland-Altman plots. Results: All test results were non-significantly different in blood samples collected with HOLDEX® single-use holder or VACUETTE® SAFELINK except potassium, whose values were marginally higher in plasma collected with VACUETTE® SAFELINK. All Pearson’s correlation coefficients were excellent except potassium, chloride and H-index. In this latter case, the correlation did not reach statistical significance. The percent bias of different parameters in samples collected with HOLDEX® single-use holder or VACUETTE® SAFELINK was minimal, comprised between −4.5% and +1.6%. In all cases the percent bias was comprised within the quality specifications tests. The rate of plasma samples with H-index >3 was eventually lower when blood was collected using SAFELINK than with HOLDEX® single use holder (odds ratio, 0.52; 95% confidence interval, 0.17–1.60). Conclusions: The results of this comparative study suggest that sample quality is thoughtfully comparable using HOLDEX® single-use holder and VACUETTE® SAFELINK, thus translating into the concept that VACUETTE® SAFELINK may also be an effective means for reducing spurious hemolysis, especially when drawing blood from catheters.

Blood sample quality using Greiner Bio-One HOLDEX® Single-Use Holder and VACUETTE® SAFELINK holder with male luer lock: a comparative study

Salvagno, Gian Luca;Pucci, Mairi;Lippi, Giuseppe
2019-01-01

Abstract

Background: Hemolysis is the most frequent and potentially one of the most harmful causes of poor sample quality. Many strategies have been attempted for minimizing the risk of generating spurious hemolysis during blood collection, such as using the HOLDEX® single-use holder, which features a chamber that decelerates blood flow pressure before entering evacuated blood tubes. We have hence designed a study to establish whether the new VACUETTE® SAFELINK, which shares different structure but similar strategy affecting blood flow pressure, may ensure a comparable sample quality as using HOLDEX® single-use holder. Methods: A total number of 24 clinical chemistry, hematologic and hemostasis parameters were measured in paired blood samples collected from 30 healthy volunteers using either HOLDEX® single-use holder or VACUETTE® SAFELINK. Test results were compared with paired Student’s t-test, Pearson’s correlation and Bland-Altman plots. Results: All test results were non-significantly different in blood samples collected with HOLDEX® single-use holder or VACUETTE® SAFELINK except potassium, whose values were marginally higher in plasma collected with VACUETTE® SAFELINK. All Pearson’s correlation coefficients were excellent except potassium, chloride and H-index. In this latter case, the correlation did not reach statistical significance. The percent bias of different parameters in samples collected with HOLDEX® single-use holder or VACUETTE® SAFELINK was minimal, comprised between −4.5% and +1.6%. In all cases the percent bias was comprised within the quality specifications tests. The rate of plasma samples with H-index >3 was eventually lower when blood was collected using SAFELINK than with HOLDEX® single use holder (odds ratio, 0.52; 95% confidence interval, 0.17–1.60). Conclusions: The results of this comparative study suggest that sample quality is thoughtfully comparable using HOLDEX® single-use holder and VACUETTE® SAFELINK, thus translating into the concept that VACUETTE® SAFELINK may also be an effective means for reducing spurious hemolysis, especially when drawing blood from catheters.
2019
blood collection, venipuncture, tube
File in questo prodotto:
File Dimensione Formato  
J Lab Precis Med 2019_06.pdf

accesso aperto

Tipologia: Documento in Post-print
Licenza: Dominio pubblico
Dimensione 163.89 kB
Formato Adobe PDF
163.89 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11562/999621
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 0
  • ???jsp.display-item.citation.isi??? ND
social impact