Automatic assessment of hemoglobin (H), lipaemia (L) and icterus (I) in serum or plasma (HIL indices) is the mainstay for evaluating sample quality. We planned this study to verify whether in-house prepared internal quality control (IQC) materials may be suitable for quality assurance of HIL indices. Pools containing different values of each of the three HIL indices were prepared from routine plasma samples, divided in aliquots and frozen at -20°C. Stability of frozen materials was assessed by thawing one aliquot of each pool after different days of freezing (1, 4, 8, 15, 22 and 29), and by measuring HIL indices on baseline fresh samples and frozen-thawed aliquots with Roche Cobas c702. Five fresh liquid IQCs materials were also measured at the same time points. Intra-assay and inter-assay imprecision of HIL indices calculated with commercial IQC materials ranged between 1.1-2.0% and 1.6-3.3%, respectively. When target values of HIL indices were calculated using frozen-thawed aliquots, the inter-assay imprecision of in-house prepared materials was optimal, even better than that of commercial liquid IQCs (H-index, 0.8% versus 1.6%; L-index, 2.2% versus 2.5%; I-index, 0.8% versus 3.3%). In conclusion, in-house prepared IQC materials are cost-effective alternatives to commercial liquid IQCs for HIL quality assurance.

Internal quality assurance of HIL indices on Roche Cobas c702

Lippi, Giuseppe;Danese, Elisa
;
Gelati, Matteo;Montagnana, Martina;Salvagno, Gian Luca;
2018-01-01

Abstract

Automatic assessment of hemoglobin (H), lipaemia (L) and icterus (I) in serum or plasma (HIL indices) is the mainstay for evaluating sample quality. We planned this study to verify whether in-house prepared internal quality control (IQC) materials may be suitable for quality assurance of HIL indices. Pools containing different values of each of the three HIL indices were prepared from routine plasma samples, divided in aliquots and frozen at -20°C. Stability of frozen materials was assessed by thawing one aliquot of each pool after different days of freezing (1, 4, 8, 15, 22 and 29), and by measuring HIL indices on baseline fresh samples and frozen-thawed aliquots with Roche Cobas c702. Five fresh liquid IQCs materials were also measured at the same time points. Intra-assay and inter-assay imprecision of HIL indices calculated with commercial IQC materials ranged between 1.1-2.0% and 1.6-3.3%, respectively. When target values of HIL indices were calculated using frozen-thawed aliquots, the inter-assay imprecision of in-house prepared materials was optimal, even better than that of commercial liquid IQCs (H-index, 0.8% versus 1.6%; L-index, 2.2% versus 2.5%; I-index, 0.8% versus 3.3%). In conclusion, in-house prepared IQC materials are cost-effective alternatives to commercial liquid IQCs for HIL quality assurance.
2018
Hemolysis, Interference, Diagnostic errors
File in questo prodotto:
File Dimensione Formato  
journal.pone.0200088.pdf

accesso aperto

Licenza: Creative commons
Dimensione 2.34 MB
Formato Adobe PDF
2.34 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11562/983256
Citazioni
  • ???jsp.display-item.citation.pmc??? 6
  • Scopus 20
  • ???jsp.display-item.citation.isi??? 17
social impact