Objectives: This study was designed to validate the analytical performance of the new Gentian particle-enhanced enhanced turbidimetric immunoassay (PETIA) for measuring neutrophil gelatinase-associated lipocalin (NGAL) in serum samples. Design and methods: Analytical validation of the Gentian NGAL assay was carried out on a Roche Cobas c501 and was based on assessment of limit of blank (LOB), limit of detection (LOD), functional sensitivity, imprecision, linearity and concordance with the BioPorto NGAL test. Results: The LOB and LOD of Gentian NGAL were found to be 3.8 ng/mL and 6.3 ng/mL, respectively. An analytical coefficient of variation (CV) of 20% corresponded to a NGAL value of 10 ng/mL. The intra-assay and inter-assay imprecision (CV) was between 0.4 and 5.2% and 0.6 and 7.1% and the total imprecision (CV) was 3.7%. The linearity was optimal at NGAL concentrations between 37 and 1420 ng/mL (r=1.00; p<0.001). An excellent correlation was observed between values measured with Gentian NGAL and BioPorto NGAL in 74 routine serum samples (r=0.993). The mean percentage bias of the Gentian assay versus the Bioporto assay was +3.1% (95% CI, +1.6% to +4.5%). Conclusions: These results show that Gentian NGAL may be a viable option to other commercial immunoassays for both routine and urgent assessment of serum NGAL.

Analytical validation of Gentian NGAL particle-enhanced enhanced turbidimetric immunoassay (PETIA)

SALVAGNO, GIAN LUCA;GELATI, Matteo;LIPPI, Giuseppe
2017-01-01

Abstract

Objectives: This study was designed to validate the analytical performance of the new Gentian particle-enhanced enhanced turbidimetric immunoassay (PETIA) for measuring neutrophil gelatinase-associated lipocalin (NGAL) in serum samples. Design and methods: Analytical validation of the Gentian NGAL assay was carried out on a Roche Cobas c501 and was based on assessment of limit of blank (LOB), limit of detection (LOD), functional sensitivity, imprecision, linearity and concordance with the BioPorto NGAL test. Results: The LOB and LOD of Gentian NGAL were found to be 3.8 ng/mL and 6.3 ng/mL, respectively. An analytical coefficient of variation (CV) of 20% corresponded to a NGAL value of 10 ng/mL. The intra-assay and inter-assay imprecision (CV) was between 0.4 and 5.2% and 0.6 and 7.1% and the total imprecision (CV) was 3.7%. The linearity was optimal at NGAL concentrations between 37 and 1420 ng/mL (r=1.00; p<0.001). An excellent correlation was observed between values measured with Gentian NGAL and BioPorto NGAL in 74 routine serum samples (r=0.993). The mean percentage bias of the Gentian assay versus the Bioporto assay was +3.1% (95% CI, +1.6% to +4.5%). Conclusions: These results show that Gentian NGAL may be a viable option to other commercial immunoassays for both routine and urgent assessment of serum NGAL.
2017
Analytical, validation, NGAL,
File in questo prodotto:
File Dimensione Formato  
Analytical validation.pdf

accesso aperto

Descrizione: Articolo principale
Tipologia: Altro materiale allegato
Licenza: Creative commons
Dimensione 283.48 kB
Formato Adobe PDF
283.48 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11562/960874
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 12
  • ???jsp.display-item.citation.isi??? 10
social impact