OBJECTIVES: Evaluation of automated flow cytometric analysis of white blood cell (WBC) count in peritoneal fluids.METHODS: One hundred peritoneal fluids were analyzed with manual microscopy, Sysmex XE-2100 and XE-5000, Siemens Advia 2120, Mindray BC-6800, Abbott Sapphire.RESULTS: High correlations (0.978 to 0.999) and modes biases (-132 to 80 WBC/mm(3)) were found. Agreement at septic peritonitis cutoff ranged between 96% and 99%.CONCLUSIONS: These hemocytometers display acceptable performance for WBC screening in peritoneal fluids.

Evaluation of white blood cell count in peritoneal fluid with five different hemocytometers.

LIPPI, Giuseppe;
2013-01-01

Abstract

OBJECTIVES: Evaluation of automated flow cytometric analysis of white blood cell (WBC) count in peritoneal fluids.METHODS: One hundred peritoneal fluids were analyzed with manual microscopy, Sysmex XE-2100 and XE-5000, Siemens Advia 2120, Mindray BC-6800, Abbott Sapphire.RESULTS: High correlations (0.978 to 0.999) and modes biases (-132 to 80 WBC/mm(3)) were found. Agreement at septic peritonitis cutoff ranged between 96% and 99%.CONCLUSIONS: These hemocytometers display acceptable performance for WBC screening in peritoneal fluids.
2013
hemocytometers; peritoneal fluid; count