The purpose of this study was to evaluate the safety and efficacy of the new 8 Fr guide catheter-compatible Flexicut directional atherectomy device and to compare it with the conventional Atherocath GTO catheter. The 6 Fr Flexicut catheter has a larger cutting window and a titanium nitride-coated cutter to effect more tissue removal as well as treat mildly calcified lesions. A group of 143 lesions in 117 consecutive patients treated with the Flexicut catheter in four centers were compared with a control group of 277 lesions in 212 consecutive patients treated with the GTO device. Postatherectomy luminal diameters were larger (2.92 +/- 0.79 vs. 2.52 +/- 0.64 mm; P < 0.0001), with more luminal gain (relative gain: 0.58 +/- 0.24 vs. 0.48 +/- 0.25; P = 0.0007) using fewer directional coronary atherectomy (DCA) cuts (12 +/- 7 vs. 16 +/- 9; P = 0.0001) in the Flexicut group. A residual diameter stenosis < 20\% immediately after DCA was obtained in 77\% of the lesions in the Flexicut group vs. 45\% in the GTO group (P < 0.0001). Histology in the former group revealed large calcium speckles in the retrieved specimens. In the Flexicut group, there was a lower incidence of access site complications and damage to the coronary ostium (2.5\% vs. 7.5\%; P = 0.08). The new Flexicut catheter is more effective than the conventional GTO catheter with a trend for reduced guiding catheter-related complications.

Effective plaque removal with a new 8 French-compatible atherectomy catheter.

RIBICHINI, Flavio Luciano;
2002-01-01

Abstract

The purpose of this study was to evaluate the safety and efficacy of the new 8 Fr guide catheter-compatible Flexicut directional atherectomy device and to compare it with the conventional Atherocath GTO catheter. The 6 Fr Flexicut catheter has a larger cutting window and a titanium nitride-coated cutter to effect more tissue removal as well as treat mildly calcified lesions. A group of 143 lesions in 117 consecutive patients treated with the Flexicut catheter in four centers were compared with a control group of 277 lesions in 212 consecutive patients treated with the GTO device. Postatherectomy luminal diameters were larger (2.92 +/- 0.79 vs. 2.52 +/- 0.64 mm; P < 0.0001), with more luminal gain (relative gain: 0.58 +/- 0.24 vs. 0.48 +/- 0.25; P = 0.0007) using fewer directional coronary atherectomy (DCA) cuts (12 +/- 7 vs. 16 +/- 9; P = 0.0001) in the Flexicut group. A residual diameter stenosis < 20\% immediately after DCA was obtained in 77\% of the lesions in the Flexicut group vs. 45\% in the GTO group (P < 0.0001). Histology in the former group revealed large calcium speckles in the retrieved specimens. In the Flexicut group, there was a lower incidence of access site complications and damage to the coronary ostium (2.5\% vs. 7.5\%; P = 0.08). The new Flexicut catheter is more effective than the conventional GTO catheter with a trend for reduced guiding catheter-related complications.
2002
Atherectomy; Coronary; instrumentation, Case-Control Studies, Catheterization, Coronary Artery Disease; pathology/surgery, Equipment Design, Female, Humans, Male, Middle Aged, Retrospective Studies
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11562/470372
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 8
  • ???jsp.display-item.citation.isi??? 6
social impact