Objective: To evaluate analytical performance of Sysmex UF-1000i for peritoneal fluid analysis. Methods: Functional sensitivity, imprecision, linearity and comparison studies were performed on peritoneal fluids. Results: Total imprecision was 1.6-4.7%, functional sensitivity 27/μL for white blood cell (WBC) and 32/μL for total nucleated cell (TNC) count. Linearity was excellent up to 983. cell/μL, carry-over < 0.2%, correlation with manual microscopy always greater than 0.992. Conclusion: The instrument exhibited optimal performance at the conventional WBC diagnostic thresholds. © 2012 The Canadian Society of Clinical Chemists.
Analytical evaluation of Sysmex UF-1000i for flow cytometric analysis of peritoneal fluid.
LIPPI, Giuseppe
2012-01-01
Abstract
Objective: To evaluate analytical performance of Sysmex UF-1000i for peritoneal fluid analysis. Methods: Functional sensitivity, imprecision, linearity and comparison studies were performed on peritoneal fluids. Results: Total imprecision was 1.6-4.7%, functional sensitivity 27/μL for white blood cell (WBC) and 32/μL for total nucleated cell (TNC) count. Linearity was excellent up to 983. cell/μL, carry-over < 0.2%, correlation with manual microscopy always greater than 0.992. Conclusion: The instrument exhibited optimal performance at the conventional WBC diagnostic thresholds. © 2012 The Canadian Society of Clinical Chemists.File in questo prodotto: