Preparation of blood specimens is a major bottleneck in the laboratory throughput. Reliable strategies for reducing the time required for specimen processing without affecting quality should be acknowledged, especially for laboratories performing stat analyses. The present investigation was planned to establish a minimal suitable centrifuge time for primary samples collected for routine coagulation testing. Five sequential primary vacuum tubes containing 0.109 mol/l buffered trisodium citrate were collected from 10 volunteers and were immediately centrifuged on a conventional centrifuge at 1500 x g, at room temperature for 1, 2, 5, 10 and 15 min, respectively. Hematological and routine coagulation testing, including prothrombin time, activated partial thromboplastin time and fibrinogen, were performed. The centrifugation time was inversely associated with residual blood cell elements in plasma, especially platelets. Statistically significant variations from the reference 15-min centrif...
Influence of the centrifuge time of primary plasma tubes on routine coagulation testing
LIPPI, Giuseppe;SALVAGNO, GIAN LUCA;MONTAGNANA, Martina;GUIDI, Giancesare
2007-01-01
Abstract
Preparation of blood specimens is a major bottleneck in the laboratory throughput. Reliable strategies for reducing the time required for specimen processing without affecting quality should be acknowledged, especially for laboratories performing stat analyses. The present investigation was planned to establish a minimal suitable centrifuge time for primary samples collected for routine coagulation testing. Five sequential primary vacuum tubes containing 0.109 mol/l buffered trisodium citrate were collected from 10 volunteers and were immediately centrifuged on a conventional centrifuge at 1500 x g, at room temperature for 1, 2, 5, 10 and 15 min, respectively. Hematological and routine coagulation testing, including prothrombin time, activated partial thromboplastin time and fibrinogen, were performed. The centrifugation time was inversely associated with residual blood cell elements in plasma, especially platelets. Statistically significant variations from the reference 15-min centrif...I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.