Objective: There is controversial evidence on a matrix influence on the measurement of NT-proBNP. Design and methods: We compared results of Elecsys NT-proBNP measurement on serum, K2 EDTA plasma and lithium heparin plasma. Results: Samples collected in K2 EDTA showed a marginally significant underestimation when compared to serum and heparin, whereas no significant difference was observed between serum and heparin plasma. Conclusions: Serum, heparin and K2 EDTA plasma may be suitable for NT-proBNP measurement. © 2007 The Canadian Society of Clinical Chemists.

Measurement of Elecsys NT-proBNP in serum, K-2 EDTA and heparin plasma

LIPPI, Giuseppe;SALVAGNO, GIAN LUCA;MONTAGNANA, Martina;GUIDI, Giancesare
2007-01-01

Abstract

Objective: There is controversial evidence on a matrix influence on the measurement of NT-proBNP. Design and methods: We compared results of Elecsys NT-proBNP measurement on serum, K2 EDTA plasma and lithium heparin plasma. Results: Samples collected in K2 EDTA showed a marginally significant underestimation when compared to serum and heparin, whereas no significant difference was observed between serum and heparin plasma. Conclusions: Serum, heparin and K2 EDTA plasma may be suitable for NT-proBNP measurement. © 2007 The Canadian Society of Clinical Chemists.
2007
EDTA; Heparin; Natriuretic peptides; Serum;
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11562/314332
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 6
  • ???jsp.display-item.citation.isi??? 5
social impact