We evaluated the analytical performances of a new commercial chemiluminescent enzyme immunometric assay for d-dimer measurement on the Immulite 2000 automated analyser. The within- and between-run coefficients of variations for low, intermediate and high d-dimer concentrations ranged from 2.2% to 5.1% and from 2.9% to 6.0% respectively. The assay was proven linear in a range of d-dimer concentrations comprised between 324 and 7602 ng/ml. Comparison of samples collected in either 0.109 mol/l sodium citrate or lithium-heparin tubes by Bland-Altman plots and nonparametric regression analysis (y = 0.858x - 39, r = 0.997; P < 0.0001) revealed a minimum bias, almost completely attributable to the dilution effect of the anticoagulant. Results of 56 outpatients' citrated plasma samples analysed with the Immulite 2000 d-dimer assay were compared with those of the current reference commercial immunoassay on the Mini Vidas Immunoanalyser. Although the nonparametric regression according to the met...
Analytical performances of the D-dimer assay for the immulite 2000 automated immunoassay analyser
LIPPI, Giuseppe;SALVAGNO, GIAN LUCA;MONTAGNANA, Martina;GUIDI, Giancesare
2007-01-01
Abstract
We evaluated the analytical performances of a new commercial chemiluminescent enzyme immunometric assay for d-dimer measurement on the Immulite 2000 automated analyser. The within- and between-run coefficients of variations for low, intermediate and high d-dimer concentrations ranged from 2.2% to 5.1% and from 2.9% to 6.0% respectively. The assay was proven linear in a range of d-dimer concentrations comprised between 324 and 7602 ng/ml. Comparison of samples collected in either 0.109 mol/l sodium citrate or lithium-heparin tubes by Bland-Altman plots and nonparametric regression analysis (y = 0.858x - 39, r = 0.997; P < 0.0001) revealed a minimum bias, almost completely attributable to the dilution effect of the anticoagulant. Results of 56 outpatients' citrated plasma samples analysed with the Immulite 2000 d-dimer assay were compared with those of the current reference commercial immunoassay on the Mini Vidas Immunoanalyser. Although the nonparametric regression according to the met...I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.