We evaluated the analytical performances of a new commercial chemiluminescent enzyme immunometric assay for d-dimer measurement on the Immulite 2000 automated analyser. The within- and between-run coefficients of variations for low, intermediate and high d-dimer concentrations ranged from 2.2% to 5.1% and from 2.9% to 6.0% respectively. The assay was proven linear in a range of d-dimer concentrations comprised between 324 and 7602 ng/ml. Comparison of samples collected in either 0.109 mol/l sodium citrate or lithium-heparin tubes by Bland-Altman plots and nonparametric regression analysis (y = 0.858x - 39, r = 0.997; P < 0.0001) revealed a minimum bias, almost completely attributable to the dilution effect of the anticoagulant. Results of 56 outpatients' citrated plasma samples analysed with the Immulite 2000 d-dimer assay were compared with those of the current reference commercial immunoassay on the Mini Vidas Immunoanalyser. Although the nonparametric regression according to the met...

Analytical performances of the D-dimer assay for the immulite 2000 automated immunoassay analyser

LIPPI, Giuseppe;SALVAGNO, GIAN LUCA;MONTAGNANA, Martina;GUIDI, Giancesare
2007-01-01

Abstract

We evaluated the analytical performances of a new commercial chemiluminescent enzyme immunometric assay for d-dimer measurement on the Immulite 2000 automated analyser. The within- and between-run coefficients of variations for low, intermediate and high d-dimer concentrations ranged from 2.2% to 5.1% and from 2.9% to 6.0% respectively. The assay was proven linear in a range of d-dimer concentrations comprised between 324 and 7602 ng/ml. Comparison of samples collected in either 0.109 mol/l sodium citrate or lithium-heparin tubes by Bland-Altman plots and nonparametric regression analysis (y = 0.858x - 39, r = 0.997; P < 0.0001) revealed a minimum bias, almost completely attributable to the dilution effect of the anticoagulant. Results of 56 outpatients' citrated plasma samples analysed with the Immulite 2000 d-dimer assay were compared with those of the current reference commercial immunoassay on the Mini Vidas Immunoanalyser. Although the nonparametric regression according to the met...
2007
D-dimer; Diagnosis; Immunoassay; Laboratory testing; Venous thromboembolism;
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11562/314298
Citazioni
  • ???jsp.display-item.citation.pmc??? 3
  • Scopus 18
  • ???jsp.display-item.citation.isi??? 16
social impact