Inertial Motor sensors (IMUs) are valid instruments for measuring postural sway but their ability to detect changes derived from visual deprivation in healthy older adults requires further investigations. We examined the validity and relationship of IMU sensor-derived postural sway measures compared to force plates for different eye conditions in healthy older adults (32 females, 33 males). We compared the relationship of the center of mass and center of pressure (CoM and CoP)-derived total length, root means square (RMS) distance, mean velocity, and 95% confidence interval ellipse area (95% CI ellipse area). In addition, we examined the relationship of the IMU sensor in discriminating between open- (EO) and closed-eye (EC) conditions compared to the force plate. A significant effect of the instruments and eye conditions was found for almost all the variables. Overall, EO and EC variables within (force plate r, from 0.38 to 0.78; IMU sensor r, from 0.36 to 0.69) as well as between (r from 0.50 to 0.88) instruments were moderately to strongly correlated. The EC:EO ratios of RMS distance and 95% CI ellipse area were not different between instruments, while there were significant differences between total length (p = 0.973) and mean velocity (p = 0.703). The ratios' correlation coefficients between instruments ranged from moderate (r = 0.65) to strong (r = 0.87). The IMU sensor offers an affordable, valid alternative to a force plate for objective, postural sway assessment.

Construct Validity of a Wearable Inertial Measurement Unit (IMU) in Measuring Postural Sway and the Effect of Visual Deprivation in Healthy Older Adults

Ferrari, Luca;Bochicchio, Gianluca;Bottari, Alberto;Pogliaghi, Silvia
2024-01-01

Abstract

Inertial Motor sensors (IMUs) are valid instruments for measuring postural sway but their ability to detect changes derived from visual deprivation in healthy older adults requires further investigations. We examined the validity and relationship of IMU sensor-derived postural sway measures compared to force plates for different eye conditions in healthy older adults (32 females, 33 males). We compared the relationship of the center of mass and center of pressure (CoM and CoP)-derived total length, root means square (RMS) distance, mean velocity, and 95% confidence interval ellipse area (95% CI ellipse area). In addition, we examined the relationship of the IMU sensor in discriminating between open- (EO) and closed-eye (EC) conditions compared to the force plate. A significant effect of the instruments and eye conditions was found for almost all the variables. Overall, EO and EC variables within (force plate r, from 0.38 to 0.78; IMU sensor r, from 0.36 to 0.69) as well as between (r from 0.50 to 0.88) instruments were moderately to strongly correlated. The EC:EO ratios of RMS distance and 95% CI ellipse area were not different between instruments, while there were significant differences between total length (p = 0.973) and mean velocity (p = 0.703). The ratios' correlation coefficients between instruments ranged from moderate (r = 0.65) to strong (r = 0.87). The IMU sensor offers an affordable, valid alternative to a force plate for objective, postural sway assessment.
2024
aging
balance
inertial sensors
postural control
risk of falls
File in questo prodotto:
File Dimensione Formato  
biosensors-14-00529.pdf

accesso aperto

Descrizione: CC BY 4.0 publisher version
Tipologia: Versione dell'editore
Licenza: Creative commons
Dimensione 1.04 MB
Formato Adobe PDF
1.04 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11562/1146368
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact