Purpose: This case report describes the possibility of XEN63 lumen obstruction at the middle of the device and emphasizes the potential to restore its flow using an ab-externo revision technique involving a 10-0 nylon suture probe, before considering more aggressive surgical interventions. Methods: A 55-year-old female with uveitic glaucoma underwent XEN63 implantation but experienced elevated intraocular pressure (IOP) (35 mm Hg) 1 month after the operation, despite medical therapy. A flat bleb, lack of response to medication, yttrium aluminum garnet (YAG) laser treatment, and slit-lamp needling necessitated revision. Results: An ab-externo surgical procedure was performed using a 10-0 nylon probe to release the obstruction, followed by the application of mitomycin C and removal of fibrous tissue and subconjunctival Tenon's capsule. Intraoperatively, flow was observed only after probing with the 10-0 nylon. Postoperatively, the patient's IOP decreased immediately and remained well-controlled at 2 months (8 mm Hg) without requiring further medication. Conclusion: For patients experiencing increased inflammatory response, the XEN63 gel stent may develop deep lumen occlusion that is unresponsive to conventional treatments. This case report introduces a novel surgical technique applicable to various glaucoma devices, utilizing an ab-externo approach with a 10-0 nylon probe. The demonstrated success in reducing IOP suggests its potential as a less invasive alternative to consider before resorting to more aggressive surgical interventions. How to cite this article: Marchese V, Randazzo V, Badalamenti R, et al. Reviving XEN63 Gel Stent Patency in Uveitic Glaucoma: A Novel Approach Using 10-0 Nylon Probe. J Curr Glaucoma Pract 2024;18(2):74-78.
Reviving XEN63 Gel Stent Patency in Uveitic Glaucoma: A Novel Approach Using 10-0 Nylon Probe
Anastasi, Marco
2024-01-01
Abstract
Purpose: This case report describes the possibility of XEN63 lumen obstruction at the middle of the device and emphasizes the potential to restore its flow using an ab-externo revision technique involving a 10-0 nylon suture probe, before considering more aggressive surgical interventions. Methods: A 55-year-old female with uveitic glaucoma underwent XEN63 implantation but experienced elevated intraocular pressure (IOP) (35 mm Hg) 1 month after the operation, despite medical therapy. A flat bleb, lack of response to medication, yttrium aluminum garnet (YAG) laser treatment, and slit-lamp needling necessitated revision. Results: An ab-externo surgical procedure was performed using a 10-0 nylon probe to release the obstruction, followed by the application of mitomycin C and removal of fibrous tissue and subconjunctival Tenon's capsule. Intraoperatively, flow was observed only after probing with the 10-0 nylon. Postoperatively, the patient's IOP decreased immediately and remained well-controlled at 2 months (8 mm Hg) without requiring further medication. Conclusion: For patients experiencing increased inflammatory response, the XEN63 gel stent may develop deep lumen occlusion that is unresponsive to conventional treatments. This case report introduces a novel surgical technique applicable to various glaucoma devices, utilizing an ab-externo approach with a 10-0 nylon probe. The demonstrated success in reducing IOP suggests its potential as a less invasive alternative to consider before resorting to more aggressive surgical interventions. How to cite this article: Marchese V, Randazzo V, Badalamenti R, et al. Reviving XEN63 Gel Stent Patency in Uveitic Glaucoma: A Novel Approach Using 10-0 Nylon Probe. J Curr Glaucoma Pract 2024;18(2):74-78.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.