The aim of this study was to verify treatment accuracy using virtual surgical planning (VSP) with a mandible-first sequence and strict surgical protocol to determine what surgical and methodological factors might influence outcomes. VSP transfer accuracy was evaluated retrospectively through a modified method involving voxelbased superimposition in patients who had undergone bimaxillary surgery with a mandible-first sequence to correct dentoskeletal deformities. Data analysis showed that the movements planned and those executed were substantially equivalent (p < 0.01), with the exception of mandibular and maxillary sagittal movements that were 0.72 ± 0.90 mm and 1.41 ± 1.04 mm smaller, respectively, than planned. This study showed that a mandible-first sequence is accurate for transferring virtual surgical planning intraoperatively. There are several factors involved in the proper transfer of virtual planning beyond the software, such as surgical technique and sequencing. Inaccurate sagittal movements and maxillary repositioning seem to depend mainly on surgical factors.

Accuracy of virtual surgical planning in bimaxillary orthognathic surgery with mandible first sequence: A retrospective study

Trevisiol, Lorenzo;Bersani, Massimo
;
D'Agostino, Antonio
2023-01-01

Abstract

The aim of this study was to verify treatment accuracy using virtual surgical planning (VSP) with a mandible-first sequence and strict surgical protocol to determine what surgical and methodological factors might influence outcomes. VSP transfer accuracy was evaluated retrospectively through a modified method involving voxelbased superimposition in patients who had undergone bimaxillary surgery with a mandible-first sequence to correct dentoskeletal deformities. Data analysis showed that the movements planned and those executed were substantially equivalent (p < 0.01), with the exception of mandibular and maxillary sagittal movements that were 0.72 ± 0.90 mm and 1.41 ± 1.04 mm smaller, respectively, than planned. This study showed that a mandible-first sequence is accurate for transferring virtual surgical planning intraoperatively. There are several factors involved in the proper transfer of virtual planning beyond the software, such as surgical technique and sequencing. Inaccurate sagittal movements and maxillary repositioning seem to depend mainly on surgical factors.
2023
orthognathic, virtual surgical planning, accuracy, mandible first
File in questo prodotto:
File Dimensione Formato  
1-s2.0-S1010518223000938.pdf

non disponibili

Descrizione: Articolo
Tipologia: Documento in Pre-print
Licenza: Copyright dell'editore
Dimensione 1.28 MB
Formato Adobe PDF
1.28 MB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11562/1097426
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 7
  • ???jsp.display-item.citation.isi??? 6
social impact