AimsInsulin pump failure and/or malfunction requiring replacement have not been thoroughly investigated. This study evaluated pump replacement in children and adolescents with Type 1 diabetes using insulin pump therapy.MethodsData were collected for all participants younger than 19 years, starting insulin pump therapy before 31 December 2013. For each child, age, disease duration, date of insulin pump therapy initiation, insulin pump model, failure/malfunction/replacement yes/no and reason were considered for the year 2013.ResultsData were returned by 40 of 43 paediatric centres belonging to the Diabetes Study Group of the Italian Society of Paediatric Endocrinology and Diabetology. In total, 1574 of 11 311 (13.9%) children and adolescents with Type 1 diabetes were using an insulin pump: 29.2% Animas VIBE, 9.4% Medtronic MiniMed 715/515, 34.3% Medtronic MiniMed VEO, 24.3% Accu-Check Spirit Combo and 2.8% other models. In 2013, 0.165 insulin pump replacements per patient-year (11.8% due to pump failure/malfunction and 4.7% due to accidental damage) were recorded. Animas VIBE (22.1%) and Medtronic MiniMed VEO (17.7%) were the most replaced.ConclusionsIn a large cohort of Italian children and adolescents with Type 1 diabetes, insulin pump failure/malfunction and consequent replacement are aligned with rates previously reported and higher in more sophisticated pump models.

Insulin pump failures in Italian children with Type 1 diabetes: retrospective 1-year cohort study

Marigliano, M.;Maffeis, C.;Sabbion, A.;
2017-01-01

Abstract

AimsInsulin pump failure and/or malfunction requiring replacement have not been thoroughly investigated. This study evaluated pump replacement in children and adolescents with Type 1 diabetes using insulin pump therapy.MethodsData were collected for all participants younger than 19 years, starting insulin pump therapy before 31 December 2013. For each child, age, disease duration, date of insulin pump therapy initiation, insulin pump model, failure/malfunction/replacement yes/no and reason were considered for the year 2013.ResultsData were returned by 40 of 43 paediatric centres belonging to the Diabetes Study Group of the Italian Society of Paediatric Endocrinology and Diabetology. In total, 1574 of 11 311 (13.9%) children and adolescents with Type 1 diabetes were using an insulin pump: 29.2% Animas VIBE, 9.4% Medtronic MiniMed 715/515, 34.3% Medtronic MiniMed VEO, 24.3% Accu-Check Spirit Combo and 2.8% other models. In 2013, 0.165 insulin pump replacements per patient-year (11.8% due to pump failure/malfunction and 4.7% due to accidental damage) were recorded. Animas VIBE (22.1%) and Medtronic MiniMed VEO (17.7%) were the most replaced.ConclusionsIn a large cohort of Italian children and adolescents with Type 1 diabetes, insulin pump failure/malfunction and consequent replacement are aligned with rates previously reported and higher in more sophisticated pump models.
2017
children and adolescents
failure
type 1 diabetes
insulin pump
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11562/1085152
Citazioni
  • ???jsp.display-item.citation.pmc??? 5
  • Scopus 14
  • ???jsp.display-item.citation.isi??? 12
social impact