This paper proposes a methodology for evaluating the effect of different stereophotogrammetric system calibration procedures on the calculation of marker-based kinematics information. The methodology, based on calibrating the system using data recorded from capture volumes of different sizes and in trials of different durations, was applied to two different systems. The calibration data were used to reconstruct the static and dynamic position and orientation in space of a rigid wand carrying markers in known positions. The inaccuracies in the reconstruction of distances and angles from the wand markers were independent on the calibration data, with average errors lower than 1.7 mm and 0.7 degrees, respectively. Similar results were obtained from human gait data, with the highest variations observed in the transverse plane kinematics and in the foot segment, suggesting that successful calibratibn procedures of different durations and performed in different volumes did not affect the metrological performance of the investigated systems.(C) 2016 The Authors. Published by Elsevier Ltd.
Effects of the calibration procedure on the metrological performances of stereophotogrammetric systems for human movement analysis
	
	
	
		
		
		
		
		
	
	
	
	
	
	
	
	
		
		
		
		
		
			
			
			
		
		
		
		
			
			
				
				
					
					
					
					
						
							
						
						
					
				
				
				
				
				
				
				
				
				
				
				
			
			
		
			
			
				
				
					
					
					
					
						
						
							
							
						
					
				
				
				
				
				
				
				
				
				
				
				
			
			
		
			
			
				
				
					
					
					
					
						
						
							
							
						
					
				
				
				
				
				
				
				
				
				
				
				
			
			
		
			
			
				
				
					
					
					
					
						
						
							
							
						
					
				
				
				
				
				
				
				
				
				
				
				
			
			
		
			
			
				
				
					
					
					
					
						
						
							
							
						
					
				
				
				
				
				
				
				
				
				
				
				
			
			
		
			
			
				
				
					
					
					
					
						
						
							
							
						
					
				
				
				
				
				
				
				
				
				
				
				
			
			
		
		
		
		
	
Di Marco, Roberto
;
	
		
		
	
			2017-01-01
Abstract
This paper proposes a methodology for evaluating the effect of different stereophotogrammetric system calibration procedures on the calculation of marker-based kinematics information. The methodology, based on calibrating the system using data recorded from capture volumes of different sizes and in trials of different durations, was applied to two different systems. The calibration data were used to reconstruct the static and dynamic position and orientation in space of a rigid wand carrying markers in known positions. The inaccuracies in the reconstruction of distances and angles from the wand markers were independent on the calibration data, with average errors lower than 1.7 mm and 0.7 degrees, respectively. Similar results were obtained from human gait data, with the highest variations observed in the transverse plane kinematics and in the foot segment, suggesting that successful calibratibn procedures of different durations and performed in different volumes did not affect the metrological performance of the investigated systems.(C) 2016 The Authors. Published by Elsevier Ltd.| File | Dimensione | Formato | |
|---|---|---|---|
| 
									
										
										
										
										
											
												
												
												    
												
											
										
									
									
										
										
											2017 DI MARCO Effects of the calibration procedure on the metrological performances of stereophotogrammetric systems for human movement analysis.pdf
										
																				
									
										
											 accesso aperto 
											Descrizione: Articolo principale
										 
									
									
									
										
											Tipologia:
											Versione dell'editore
										 
									
									
									
									
										
											Licenza:
											
											
												Creative commons
												
												
													
													
													
												
												
											
										 
									
									
										Dimensione
										533.57 kB
									 
									
										Formato
										Adobe PDF
									 
										
										
								 | 
								533.57 kB | Adobe PDF | Visualizza/Apri | 
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.