Introduction: Point of care testing (POCT) represents a valuable option when laboratory data shall be urgently available for timely clinical management, with a turnaround time (TAT) that is unfeasible using conventional laboratory instrumentation. This study was aimed to compare the performance of QBC STAR™ compared to Sysmex XN-module and the reference optical microscopy (OM) assessment. Material and methods: One hundred peripheral blood samples, collected in K3 EDTA tubes, and 50 capillary blood samples obtained by finger stick were analyzed with QBC STAR™, Sysmex XN-module, and OM. Data were compared with Passing-Bablok regression and Bland-Altman plots. Results: The Passing-Bablok regression analysis (QBC STAR™ capillary sample vs XN-module) yielded slopes comprised between 0.30 and 1.37, while the intercepts ranged between -17.57 and 232.6. Bland-Altman plots yielded relative bias comprised between -4.87% (for MN QBC STAR™ capillary sample vs XN-module) and 27% (PLT QBC STAR™ capillary sample vs XN-module). A significant bias was found for all parameters except MN and WBC, RBC in all and pediatric samples, and HB in adults samples. Conclusion: The results of this analytical evaluation suggest that QBC STAR™ may not be the ideal tool for performing complete blood count analysis for diagnostic purposes, while it could be more useful in urgent/emergent conditions, such as for rapid monitoring of some hematological parameters (eg, WBC and HB).

Complete Blood Count as point of care testing QBC STAR™: Preliminary evaluation

Lippi, Giuseppe;
2021-01-01

Abstract

Introduction: Point of care testing (POCT) represents a valuable option when laboratory data shall be urgently available for timely clinical management, with a turnaround time (TAT) that is unfeasible using conventional laboratory instrumentation. This study was aimed to compare the performance of QBC STAR™ compared to Sysmex XN-module and the reference optical microscopy (OM) assessment. Material and methods: One hundred peripheral blood samples, collected in K3 EDTA tubes, and 50 capillary blood samples obtained by finger stick were analyzed with QBC STAR™, Sysmex XN-module, and OM. Data were compared with Passing-Bablok regression and Bland-Altman plots. Results: The Passing-Bablok regression analysis (QBC STAR™ capillary sample vs XN-module) yielded slopes comprised between 0.30 and 1.37, while the intercepts ranged between -17.57 and 232.6. Bland-Altman plots yielded relative bias comprised between -4.87% (for MN QBC STAR™ capillary sample vs XN-module) and 27% (PLT QBC STAR™ capillary sample vs XN-module). A significant bias was found for all parameters except MN and WBC, RBC in all and pediatric samples, and HB in adults samples. Conclusion: The results of this analytical evaluation suggest that QBC STAR™ may not be the ideal tool for performing complete blood count analysis for diagnostic purposes, while it could be more useful in urgent/emergent conditions, such as for rapid monitoring of some hematological parameters (eg, WBC and HB).
2021
point of care testing
QBC STAR™, Sysmex XN, complete blood count, point of care testing
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11562/1040142
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 0
  • ???jsp.display-item.citation.isi??? 1
social impact