Backgrounds Central arterial pressure can be derived from analysis of the peripheral artery waveform. The aim of this study was to compare central arterial pressures measured from an intra-aortic catheter with peripheral radial arterial pressures and with central arterial pressures estimated from the peripheral pressure wave using a pressure recording analytical method (PRAM). Methods We studied 21 patients undergoing digital subtraction cerebral angiography under local or general anesthesia and equipped with a radial arterial catheter. A second catheter was placed in the ascending aorta for central pressure wave acquisition. Central (AO) and peripheral (RA) arterial waveforms were recorded simultaneously by PRAM for 90-180 s. During an off-line analysis, AO pressures were reconstructed (AO(rec)) from the RA trace using a mathematical model obtained by multi-linear regression analysis. The AO(rec) values obtained by PRAM were compared with the true central pressure value obtained from the catheter placed in the ascending aorta. Results Systolic, diastolic and mean pressures ranged from 79 to 180 mmHg, 47 to 102 mmHg, and 58 to 128 mmHg, respectively, for AO, and 83 to 174 mmHg, 47 to 107 mmHg, and 60 to 129 mmHg, respectively, for RA. The correlation coefficients between AO and RA were 0.86 (p < 0.01), 0.83 (p < 0.01) and 0.86 (p < 0.01) for systolic, diastolic and mean pressures, respectively, and the mean differences - 0.3 mmHg, 2.4 mmHg and 1.5 mmHg. The correlation coefficients between AO and AO(rec) were 0.92 (p < 0.001), 0.87 (p < 0.001) and 0.92 (p < 0.001), for systolic, diastolic and mean pressures, respectively, and the mean differences 0.01 mmHg, 1.8 mmHg and 1.2 mmHg. Conclusions PRAM can provide reliable estimates of central arterial pressure.

Estimation of central arterial pressure from the radial artery in patients undergoing invasive neuroradiological procedures

Donadello, Katia
Writing – Review & Editing
;
Franchi, Federico;Polati, Enrico
Writing – Review & Editing
;
2019-01-01

Abstract

Backgrounds Central arterial pressure can be derived from analysis of the peripheral artery waveform. The aim of this study was to compare central arterial pressures measured from an intra-aortic catheter with peripheral radial arterial pressures and with central arterial pressures estimated from the peripheral pressure wave using a pressure recording analytical method (PRAM). Methods We studied 21 patients undergoing digital subtraction cerebral angiography under local or general anesthesia and equipped with a radial arterial catheter. A second catheter was placed in the ascending aorta for central pressure wave acquisition. Central (AO) and peripheral (RA) arterial waveforms were recorded simultaneously by PRAM for 90-180 s. During an off-line analysis, AO pressures were reconstructed (AO(rec)) from the RA trace using a mathematical model obtained by multi-linear regression analysis. The AO(rec) values obtained by PRAM were compared with the true central pressure value obtained from the catheter placed in the ascending aorta. Results Systolic, diastolic and mean pressures ranged from 79 to 180 mmHg, 47 to 102 mmHg, and 58 to 128 mmHg, respectively, for AO, and 83 to 174 mmHg, 47 to 107 mmHg, and 60 to 129 mmHg, respectively, for RA. The correlation coefficients between AO and RA were 0.86 (p < 0.01), 0.83 (p < 0.01) and 0.86 (p < 0.01) for systolic, diastolic and mean pressures, respectively, and the mean differences - 0.3 mmHg, 2.4 mmHg and 1.5 mmHg. The correlation coefficients between AO and AO(rec) were 0.92 (p < 0.001), 0.87 (p < 0.001) and 0.92 (p < 0.001), for systolic, diastolic and mean pressures, respectively, and the mean differences 0.01 mmHg, 1.8 mmHg and 1.2 mmHg. Conclusions PRAM can provide reliable estimates of central arterial pressure.
2019
Humans
Models, Theoretical
Arterial pressure
Central pressure waveform
Peripheral artery waveform
Pulse contour analysis
Adult
Angiography, Digital Subtraction
Aorta
Arterial Pressure
Blood Pressure
Cerebral Angiography
File in questo prodotto:
File Dimensione Formato  
2019_Scolletta_BMC_Anesthesiology.pdf

accesso aperto

Tipologia: Documento in Post-print
Licenza: Dominio pubblico
Dimensione 806.61 kB
Formato Adobe PDF
806.61 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11562/1037189
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 4
  • ???jsp.display-item.citation.isi??? 6
social impact