Background Filling of citrate tubes with appropriate amount of blood is essential for obtaining reliable results of coagulation testing. This study aimed to verify whether results of routine coagulation tests are comparable when the new Becton Dickinson Vacutainer® Citrate Plus tubes are filled at minimum or optimal volume. Methods The study population consisted of 133 patients (40 on oral anticoagulant therapy), who had blood collected for routine coagulation testing. Two sequential Vacutainer® Citrate Plus tubes of the same type and lot were drawn. The first tube was collected after a butterfly needle was inserted into the vein, so that the air in the tubing was aspirated into the tube before blood (minimum fill volume), whilst the second was drawn at optimal fill volume. Experiments were repeated using 2.7-mL (n = 86) and 1.8-mL (n = 47) tubes. Results Prothrombin time (PT) and fibrinogen values were slightly but significantly decreased in tubes with minimum than in those with optimal fill volume. The activated partial thromboplastin time (APTT) was slightly prolonged in tubes with minimum than in those with optimal fill volume, but the difference was not statistically significant. An identical trend was noted in separate analyses for the 2.7-mL and 1.8-mL tubes. Spearman's correlations between the two fill volumes were always >0.94 and bias was always within the quality specifications. Conclusions Blood drawing into Vacutainer® Citrate Plus tubes at minimum fill volume does not clinically bias routine coagulation testing.

Routine coagulation testing in Vacutainer® Citrate Plus tubes filled at minimum or optimal volume

Lippi, Giuseppe
2020-01-01

Abstract

Background Filling of citrate tubes with appropriate amount of blood is essential for obtaining reliable results of coagulation testing. This study aimed to verify whether results of routine coagulation tests are comparable when the new Becton Dickinson Vacutainer® Citrate Plus tubes are filled at minimum or optimal volume. Methods The study population consisted of 133 patients (40 on oral anticoagulant therapy), who had blood collected for routine coagulation testing. Two sequential Vacutainer® Citrate Plus tubes of the same type and lot were drawn. The first tube was collected after a butterfly needle was inserted into the vein, so that the air in the tubing was aspirated into the tube before blood (minimum fill volume), whilst the second was drawn at optimal fill volume. Experiments were repeated using 2.7-mL (n = 86) and 1.8-mL (n = 47) tubes. Results Prothrombin time (PT) and fibrinogen values were slightly but significantly decreased in tubes with minimum than in those with optimal fill volume. The activated partial thromboplastin time (APTT) was slightly prolonged in tubes with minimum than in those with optimal fill volume, but the difference was not statistically significant. An identical trend was noted in separate analyses for the 2.7-mL and 1.8-mL tubes. Spearman's correlations between the two fill volumes were always >0.94 and bias was always within the quality specifications. Conclusions Blood drawing into Vacutainer® Citrate Plus tubes at minimum fill volume does not clinically bias routine coagulation testing.
2020
anticoagulant; blood tubes; coagulation; hemostasis
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11562/1010116
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 0
  • ???jsp.display-item.citation.isi??? 0
social impact