This study assesses the use of different dry K2 (dipotassium) EDTA vacuum tubes and whether or not they might represent a bias in haematological testing. Blood was collected in three dipotassium EDTA vacuum tubes from different manufacturers: Venosafe, Vacuette and Vacutainer. Samples were analysed on an Advia 2120i analyser. Significant differences among results and biases were compared with current quality specifications. Significant differences were found for haematocrit (HCT), mean corpuscular volume (MCV), white blood cell count (WBC) and platelet distribution width (PDW) when comparing Venosafe vs. Vacuette; for MCV, WBC and PDW when comparing Venosafe vs. Vacutainer; and for HCT and MCV when comparing Vacuette vs. Vacutainer. Clinically significant variations were observed for HCT and PDW in Venosafe vs. Vacuette; PDW in Venosafe vs. Vacutainer; and HCT and MCV in Vacuette vs. Vacutainer. The use of dipotassium EDTA vacuum tubes from different manufacturers represent a clinically relevant source of variation for HCT, MCV and PDW.

Brand of dipotassium EDTA vacuum tube as a new source of pre-analytical variability in routine haematology testing.

De Souza Lima Oliveira, Gabriel
;
LIPPI, Giuseppe;SALVAGNO, GIAN LUCA;MONTAGNANA, Martina;Poli, Giovanni;GUIDI, Giancesare
2013-01-01

Abstract

This study assesses the use of different dry K2 (dipotassium) EDTA vacuum tubes and whether or not they might represent a bias in haematological testing. Blood was collected in three dipotassium EDTA vacuum tubes from different manufacturers: Venosafe, Vacuette and Vacutainer. Samples were analysed on an Advia 2120i analyser. Significant differences among results and biases were compared with current quality specifications. Significant differences were found for haematocrit (HCT), mean corpuscular volume (MCV), white blood cell count (WBC) and platelet distribution width (PDW) when comparing Venosafe vs. Vacuette; for MCV, WBC and PDW when comparing Venosafe vs. Vacutainer; and for HCT and MCV when comparing Vacuette vs. Vacutainer. Clinically significant variations were observed for HCT and PDW in Venosafe vs. Vacuette; PDW in Venosafe vs. Vacutainer; and HCT and MCV in Vacuette vs. Vacutainer. The use of dipotassium EDTA vacuum tubes from different manufacturers represent a clinically relevant source of variation for HCT, MCV and PDW.
2013
Anticoagulants; Blood cell count; Potassium EDTA; Pre-analytical phase
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11562/560150
Citazioni
  • ???jsp.display-item.citation.pmc??? 6
  • Scopus 18
  • ???jsp.display-item.citation.isi??? 15
social impact